What is a toxicological risk assessment for a medical device?
It is the assessment of whether the chemical constituents a device can release could harm a patient at the amounts measured. Under ISO 10993-17:2023, each constituent of concern is given a tolerable intake, the worst-case patient exposure is estimated, and the two are compared as a margin of safety.
What is ISO 10993-17?
ISO 10993-17 is the part of the ISO 10993 series that covers the toxicological risk assessment of medical device constituents. The current edition, ISO 10993-17:2023, sets out how to screen constituents, derive tolerable intakes, estimate exposure and calculate a margin of safety.
Is a toxicological risk assessment a test?
No. It is a documented assessment, not a laboratory test. It uses the results of chemical characterization testing under ISO 10993-18, together with published and predicted toxicity data.
What do you need from us to start?
A chemical characterization report that identifies and quantifies the extractable or leachable constituents, a description of the device, its contact type and duration, the number of devices a patient may be exposed to, and the intended patient population.
Can a toxicological risk assessment replace biocompatibility testing?
It can address some endpoints. ISO 10993-1 allows chemical characterization with a toxicological risk assessment to be used as evidence for endpoints such as systemic toxicity, genotoxicity and carcinogenicity, where the data support it. Whether that evidence is sufficient for a given device is decided in the biological evaluation.
What is the toxicological screening limit?
The toxicological screening limit (TSL) is a cumulative exposure below which a constituent is not expected to pose a toxicological concern and does not need an individual assessment. Constituents above it are carried forward.
What happens if a margin of safety is below 1?
A margin of safety below 1 means the worst-case estimate exceeds the tolerable intake, so the constituent needs further work. That can mean refining the exposure estimate with more realistic data, further chemistry, a targeted biological test, or a change to the material or process.
Who carries out the assessment?
Every Groenakker toxicological risk assessment is reviewed by a board-certified toxicologist (DABT).
How long does a toxicological risk assessment take?
Within a ChemTox™ study, data analysis and toxicological review take an estimated one to two weeks once the chemistry is complete, and reporting about a further week. A standalone assessment on chemistry data you already hold takes an estimated two to three weeks from receipt of complete data.