At a glance

You receive
A constituent-by-constituent assessment in a DABT-reviewed report
Turnaround
2 to 3 weeks from complete chemistry data (estimate)
Standard
ISO 10993-17:2023
We need from you
Chemical characterization report, device description, contact type and duration

What is a toxicological risk assessment?

A toxicological risk assessment (TRA) establishes whether the chemical constituents a medical device can release are present at levels that could harm a patient. It is carried out under ISO 10993-17:2023 and takes the results of chemical characterization, the extractables and leachables testing described in ISO 10993-18, as its starting point.

Chemistry shows what is there and how much; the assessment shows whether that amount matters. For each constituent above the toxicological screening limit, a tolerable intake is derived from the available toxicity data, a worst-case exposure is estimated from the amount measured and the way the device is used, and the two are compared as a margin of safety.

The conclusions feed the device’s biocompatibility evaluation, where they can be used as evidence for endpoints such as systemic toxicity, genotoxicity and carcinogenicity. Groenakker carries out the assessment on its own or together with the chemistry as a ChemTox™ extractables and leachables study.

What you receive

Every value traceable to its source

A toxicological risk assessment is only as sound as the chemistry behind it and the reasoning recorded in it. Each tolerable intake in a Groenakker report is shown with the study it came from and the factors applied, so a reviewer can follow the value back to its source.

The conclusions are written to be cited. Each one names the endpoint it addresses, so your biological evaluation report can reference it directly, and every assessment is reviewed by a board-certified toxicologist (DABT).

Included

  • Constituent-by-constituent assessmentTolerable intake, worst-case exposure and margin of safety for each constituent carried forward.
  • Screening recordConstituents compared with the toxicological screening limit, EU MDR CMR substances and cohort-of-concern compounds.
  • Endpoint conclusionsA conclusion for each toxicological endpoint the assessment addresses.
  • DABT-reviewed reportIssued under Groenakker cover, ready to reference from your biological evaluation report.

How it works

Within a ChemTox™ study the data analysis and toxicological review take an estimated one to two weeks, and reporting about a further week. A standalone assessment takes an estimated two to three weeks from receipt of complete chemistry data.

Phase I

Scope and data review

01 – Agree the scope: the device, its contact type and duration, and the chemistry data to be assessed.

02 – Groenakker reviews the chemical characterization report and confirms it identifies and quantifies constituents to the level the assessment needs.

Phase II

Screening

03 – Reported constituents are compared with the toxicological screening limit; those above it are carried forward.

04 – Constituents are screened against EU MDR CMR substances and cohort-of-concern compounds.

Phase III

Assessment

05ᵃ – Toxicity data are gathered and a tolerable intake is derived for each constituent carried forward.

05ᵇ – Worst-case exposure is estimated and a margin of safety is calculated for each constituent.

Phase IV

Reporting

06 – The report is reviewed by a board-certified toxicologist (DABT) and issued to the client.

The standalone assessment is listed in the test catalog, where you can add it to an estimate.

Three ways to commission an assessment

An assessment can be commissioned in three ways, depending on what chemistry data you already hold.

Standalone assessment

On chemistry data you already hold

If your device has already been through chemical characterization, the assessment can be carried out on that report. Groenakker reviews the data first and tells you if anything is missing before the assessment starts.

What is included

  • Review of the existing ISO 10993-18 chemical characterization report
  • A list of gaps before work starts, if identification or quantification falls short
  • Assessment and report to ISO 10993-17:2023
In depth

How an ISO 10993-17:2023 assessment is built

The 2023 edition of ISO 10993-17 sets out the assessment as a sequence. Constituents reported by the chemistry are first compared with the toxicological screening limit. Those above it are assessed one by one, and each assessment ends in a margin of safety that can be traced back to the data it came from.

Tolerable intake (TI)

The amount of a constituent a patient can be exposed to without appreciable harm.

  • Derived for each constituent carried forward
  • Starts from a point of departure in the toxicity data, such as a no-observed-adverse-effect level
  • Adjusted with documented uncertainty factors
  • The basis for each value is recorded in the report

Worst-case estimated exposure

How much of the constituent a patient could receive from the device.

  • Based on the quantity of the constituent extracted from the device
  • Scaled for the number of devices used and the duration of contact
  • Expressed for the body weight of the intended patient population
  • Assumptions chosen to overestimate exposure rather than underestimate it

Margin of safety (MoS)

The comparison that gives the conclusion for each constituent.

  • The tolerable intake divided by the worst-case estimated exposure
  • Greater than 1 means the estimated exposure is below the tolerable intake
  • Calculated constituent by constituent
  • Where a margin is not met, the report sets out what further evidence is needed

Constituents with little or no data

Not every extractable has a published toxicity study. Where data are missing, the assessment draws on:

  • Structure-based (QSAR) and other in silico predictions
  • Grouping with closely related compounds
  • Screening against EU MDR CMR substances and cohort-of-concern compounds

Toxicological risk assessment FAQ

What is a toxicological risk assessment for a medical device?

It is the assessment of whether the chemical constituents a device can release could harm a patient at the amounts measured. Under ISO 10993-17:2023, each constituent of concern is given a tolerable intake, the worst-case patient exposure is estimated, and the two are compared as a margin of safety.

What is ISO 10993-17?

ISO 10993-17 is the part of the ISO 10993 series that covers the toxicological risk assessment of medical device constituents. The current edition, ISO 10993-17:2023, sets out how to screen constituents, derive tolerable intakes, estimate exposure and calculate a margin of safety.

Is a toxicological risk assessment a test?

No. It is a documented assessment, not a laboratory test. It uses the results of chemical characterization testing under ISO 10993-18, together with published and predicted toxicity data.

What do you need from us to start?

A chemical characterization report that identifies and quantifies the extractable or leachable constituents, a description of the device, its contact type and duration, the number of devices a patient may be exposed to, and the intended patient population.

Can a toxicological risk assessment replace biocompatibility testing?

It can address some endpoints. ISO 10993-1 allows chemical characterization with a toxicological risk assessment to be used as evidence for endpoints such as systemic toxicity, genotoxicity and carcinogenicity, where the data support it. Whether that evidence is sufficient for a given device is decided in the biological evaluation.

What is the toxicological screening limit?

The toxicological screening limit (TSL) is a cumulative exposure below which a constituent is not expected to pose a toxicological concern and does not need an individual assessment. Constituents above it are carried forward.

What happens if a margin of safety is below 1?

A margin of safety below 1 means the worst-case estimate exceeds the tolerable intake, so the constituent needs further work. That can mean refining the exposure estimate with more realistic data, further chemistry, a targeted biological test, or a change to the material or process.

Who carries out the assessment?

Every Groenakker toxicological risk assessment is reviewed by a board-certified toxicologist (DABT).

How long does a toxicological risk assessment take?

Within a ChemTox™ study, data analysis and toxicological review take an estimated one to two weeks once the chemistry is complete, and reporting about a further week. A standalone assessment on chemistry data you already hold takes an estimated two to three weeks from receipt of complete data.