When Biocompatibility Tests Fail: How to Find the Root Cause

When a biocompatibility test fails unexpectedly, the lab is rarely to blame. Here is how to find the root cause, and why chemistry is the one specific tool. If you manage biocompatibility testing for any length of time, you will very likely face an unexpected failure, usually on the project that can least afford it. This walks through what the results and the lab’s extract observations can still tell you, which interim tests are worth the time, and why extractables testing under ISO 10993-18 is usually the fastest route to a definitive answer.

As is the way of things, it will probably happen on a critical project that is already behind schedule, where biocompatibility testing is the only thing holding up submission and launch. The one thing you could not afford to go wrong has gone wrong. What do you do?

The first inclination, whether you like to admit it or not, is to find a way, directly or quietly, to blame the lab. A lab-based issue is a relatively tidy way to justify a retest and move on. But it is most often not the lab that caused the failure. Test controls and validated methods go a long way toward ensuring that, and if the negative, positive and blank controls all performed as expected, the assay itself is usually not the problem.

So once the lab blame game is over, what do you really do?

This is where next steps get tricky, time-consuming and expensive. Assume for a moment that the test articles were produced correctly, and so represent the saleable condition of the product, and that the selected test and conditions were appropriate for the device and its intended use. Then there is a legitimate concern that the device may not, in fact, be fully biocompatible.

But why?

What the test can still tell you

The unfortunate truth about most biocompatibility tests is that they give a clear indication of pass or fail but do a very poor job of pointing to why.

That does not mean they offer no clues. If it is an extract-based test, a good first step is to review the lab's notes and observations on the extract:

  • Was there a change in color or pH?

  • Did the extract turn cloudy?

  • Did the device swell or show any signs of degradation?

  • Which extraction vehicle did the failure occur in: polar, non-polar or both?

While you are asking, ask the lab to hold any remaining extract. Labs often discard it once the test is read, and if it is still available it can be analyzed directly, which ties the chemistry to the exact sample that failed.

The answers can point you in a direction and narrow the field of inquiry. If you have done your due diligence and built the device from well-known biocompatible materials, they can also help narrow down which of your processes may not have performed as expected:

  • Residual manufacturing materials, such as mold release or machining fluids?

  • Cleaning detergent residue?

  • Silicone or adhesive that was not fully or properly cured?

When the answers do not come easily

In a perfect world, the answers come quickly, and the extract observations and manufacturing records point to a clear smoking gun. Reality is usually less straightforward, and it tends to bring a period of significant churn: endless calls and meetings with the lab and the manufacturer, and hours spent digging through manufacturing records and test data for a direction.

Depending on the nature of the failure, someone may recommend a dilution series or other interim tests in the hope of a pointer to the root cause. Be selective here. Most interim testing costs weeks and money and ends in a recommendation to run chemistry anyway, or leaves you holding a result and asking, “OK, so what?”

The interim tests worth the time share one property: the result changes what you do next.

Tests that can exonerate the device

Measuring the pH and osmolality of the extract is neither slow nor especially expensive, and the cell monolayer has no defense against either. An extract that has drifted osmotically will kill cells without any meaningful leachable being involved at all. If that turns out to be what happened, the finding reframes the whole investigation: the device did not fail; the extract did.

Tests that check a named suspect

If you have a hypothesis, such as a resin change, a new supplier or an adhesive you are not certain cured, FTIR against a known-good article will generally confirm or eliminate it quickly enough to be worth doing.quickly enough to be worth doing. Bear in mind that FTIR identifies bulk and surface composition; a trace residue can sit below what it detects, so a clean spectrum narrows the field rather than closing it. Similarly, if the device is ethylene oxide sterilized, EO and ECH residuals under ISO 10993-7:2008 (and its 2019 amendment) are a common enough cause to rule out early rather than late.

What is rarely worth doing is running a panel of interim assays in the hope that something turns up. That is, more often than not, where the weeks and the money go. Meanwhile you are spending as much of your day preparing updates for management on why the submission is on hold as you are solving the problem.

So the question comes back: if this is your reality, what do you do?

Why chemistry is the one specific tool

When you have reviewed the data and come up empty, and assuming chemical analysis has not already been done and informed your review, get moving on extractables testing. That means chemical characterization under ISO 10993-18:2020, followed by a toxicological risk assessment under ISO 10993-17:2023.

A chemical panel is the one part of the biocompatibility suite that delivers discrete results: you have this much of this constituent, and it was extracted from your product under the test conditions. If you want to understand what another test reacted to, a well-performed chemical analysis is highly likely to find it. That includes inorganics: if metal ions such as nickel, copper or silver are plausible, make sure the panel includes an elemental analysis (typically ICP-MS) alongside the organic screens.

Chemistry was probably already somewhere in your testing plan. Do not put it on hold; move it forward. Chemistry and toxicology are not cheap, but chemistry is the one tool you have that is specific. Every other step in an investigation leads either to another test or to a hoped-for rationale. In nearly every case where we have used it, chemistry has led to a definitive answer, and it often supplies the detail needed both to explain the failure and to support the device's safety for its intended use.

Scope it with the failure in mind. Extractables studies normally use exaggerated or exhaustive extractions in polar, semi-polar and non-polar solvents, which is right for a safety assessment but not identical to the culture-medium extract that failed. Where the goal is explaining a specific result, add an extraction that mirrors the failed test, or analyze the retained extract, so the findings can be tied back to it.

How much the science has changed

This was not always the case. As little as ten years ago, submitting a device for a general extractables screen was the equivalent of throwing stacks of cash into a burning dumpster, the only difference being that the dumpster at least produced something useful in the form of heat.

That is no longer true. The industry as a whole has done an immense amount of work to bring the science to a point where it delivers meaningful, actionable data that is, above all, trustworthy. ISO 10993-18 was rewritten in 2020 and amended in 2022, ISO 10993-17 followed in 2023 with a full method for toxicological risk assessment, and FDA issued draft guidance on chemical analysis for biocompatibility in September 2024.

What chemistry will not do

Is chemistry the right answer in every situation? No.

Will chemistry always give you everything you need and solve your problem? No.

Results come with limits of their own. Identifications carry confidence levels, many concentrations are semi-quantitative estimates adjusted with an uncertainty factor, and it is the toxicological risk assessment, not the chemistry alone, that decides whether a constituent matters at the level found.

There are no absolute guarantees. Your data may tell you the problem is bigger than you thought, and that the device has a fundamental issue. Even so, it is rare for chemistry data not to help set a path forward.

Do the other steps first and investigate properly. Make sure none of the basics were missed. But if they come up empty, do not forget chemistry. It is likely the difference between directionless churn and a definitive path forward.

Working through a failure now? Talk to a specialist. We scope the extractables study and toxicological risk assessment around the result you are trying to explain, so the chemistry answers your question rather than a generic one.


Part 1 of this series: Cytotoxicity Testing: The Test Most Likely to Fail, on how the test works, why its simplicity makes it so sensitive, and the three sampling principles that make any result one you can trust.