EN ISO 10993-6:2016 and EN ISO 10993-7:2008/A1:2019 cited without a presumption of conformity (never on the MDR list)
EN ISO 10993-6:2026 and EN ISO 10993-7:2026 — harmonised under Regulation (EU) 2017/745 by Decision (EU) 2026/2479, OJ L 10 November 2026. US files unchanged: ISO 10993-6:2016 (2-247), ISO 10993-7:2008 (2-275)
Decision (EU) 2026/2479 lists the 2026 editions of Parts 6 and 7 as harmonised under the MDR for the first time. FDA still recognizes the 2016 and 2008 editions.
- Who it affects
- EU MDR technical documentation; implantation and EO residual data in EU files; IVDR files using radiation sterilization
What changed
Commission Implementing Decision (EU) 2026/2479 of 9 November 2026 was published in the Official Journal on 10 November 2026 and entered into force the same day. It adds EN ISO 10993-6:2026 (tests for local effects after implantation) and EN ISO 10993-7:2026 (ethylene oxide sterilization residuals) to the list of standards harmonised under Regulation (EU) 2017/745. Neither part had been on the MDR list before, so these are new listings rather than replacements, and no earlier edition is being withdrawn.
The same Decision lists EN ISO 11137-1:2026 (radiation sterilization) and EN ISO 14155:2026 (clinical investigations), and deletes the 2015 and 2020 references from 10 May 2028. A companion Decision, (EU) 2026/2481, adds EN ISO 11137-1:2026 to the IVDR list with the same 2028 withdrawal.
For US files nothing has moved. As of 20 November 2026, FDA recognizes ISO 10993-6:2016 (recognition 2-247, complete) and ISO 10993-7:2008 with its 2009 corrigendum and 2019 amendment (recognition 2-275, complete), and has no entry for either 2026 edition.
Who it affects
Manufacturers relying on the presumption of conformity for implantation testing or EO residual limits in EU technical documentation, and anyone running a single biological evaluation for both EU and US markets.
What to do
In EU files, cite the 2026 editions where you rely on the presumption of conformity. Where implantation or EO residual data were generated to the 2016 or 2008 editions, add a short bridging rationale that maps the method and acceptance criteria to the 2026 text; most data carry over, but the rationale should say so explicitly. In US files, keep citing the FDA-recognized editions and state the recognition number. For dual-market programs, write the BEP so that one dataset is mapped to both references rather than running two plans.
Sources
- Commission Implementing Decision (EU) 2026/2479 checked 20 Nov 2026
- Commission Implementing Decision (EU) 2026/2481 checked 20 Nov 2026
- FDA recognition 2-247, ISO 10993-6:2016 checked 20 Nov 2026
- FDA recognition 2-275, ISO 10993-7:2008 checked 20 Nov 2026