ISO 10993-18:2020/Amd 1:2022 with the Annex E uncertainty factor of 2 for GC screening and Table D.3
ISO 10993-18:2020 + Amd 1:2022 — FDA recognition 2-298 (partial: clause 5.5 sentences 2–3, the clause 7 TRA phrase, Table D.3 and the Annex E UF=2 and multi-method sentences are not recognized); one justified uncertainty factor per analytical method
FDA's recognition 2-298 of ISO 10993-18:2020 with Amd 1:2022 excludes Table D.3 and the Annex E text that offers an uncertainty factor of 2 for GC screening. Each method needs its own justified factor.
- Who it affects
- Chemical characterization reports and AET derivations in US submissions
What changed
FDA entered recognition 2-298 for ISO 10993-18:2020 including Amendment 1:2022 on 7 September 2026. The recognition is partial. Not recognized: the second and third sentences of clause 5.5; the phrase "to assist in any toxicological risk assessment" in the second paragraph of clause 7; Table D.3 in Annex D; the Annex E sentence that proposes an uncertainty factor of 2 as appropriate for screening semi-volatile extractables; and the last two sentences of the Annex E paragraph that allow one uncertainty factor to cover several complementary analytical methods.
FDA's rationale on the uncertainty factor cites the Jenke and Odufu (2012) publication that the standard itself relies on, and reads it as showing that an analytical evaluation threshold covering 90% of the response-factor population needs a factor of about 4, not 2. On multiple methods, FDA cites Sussman et al. (2022): the uncertainty factor depends on the analytical method, and low-response-factor analytes seen on a single system are the ones most likely to be under-reported.
FDA's "Chemical Analysis for Biocompatibility Assessment of Medical Devices" (September 2024, docket FDA-2024-D-4165) remains draft guidance and is marked not for implementation.
Who it affects
Any US submission with an ISO 10993-18 chemistry package, and in particular any AET derived with a generic uncertainty factor of 2 or a single factor applied across GC-MS, LC-MS and ICP-MS.
What to do
Justify the uncertainty factor for each analytical method from that method's own response-factor data, and state it in the report. Do not cite Table D.3 as a recognized source. Where an existing report used a factor of 2 for GC screening, expect a deficiency and prepare the re-derivation before submission rather than after. For EU files the standard applies as published, but a method-specific factor is defensible in both markets.
Sources
- FDA recognition 2-298, ISO 10993-18:2020/Amd 1:2022 checked 20 Nov 2026
- FDA draft guidance, Chemical Analysis for Biocompatibility Assessment of Medical Devices checked 20 Nov 2026