All regulatory updates

Regulatory updateStandards and recognition

FDA recognizes ISO 10993-1:2025; 2018 declarations accepted until 1 July 2029

  • US United States (FDA)
Before

ISO 10993-1:2018 (FDA recognition 2-258); accepted until 1 July 2029

Cite now

ISO 10993-1:2025, Sixth edition 2025-11 — FDA recognition 2-313 (partial: clause 6.9 and "consumer products" in clause 6.5.11.3 are not recognized); ISO 10993-12:2021 + Amd 1:2025 — recognition 2-314

Recognition List 066 replaced 2-258 (2018 edition) with 2-313 (2025 edition, partial). The 2018 edition stays acceptable until 1 July 2029, and FDA's September 2023 guidance is unchanged.

Who it affects
Every US submission that declares conformity to ISO 10993-1; biological evaluation plans in progress

What changed

Recognition List Number 066, published in the Federal Register on 24 August 2026 (91 FR 54715), withdrew recognition 2-258 for ISO 10993-1:2018 and replaced it with 2-313 for ISO 10993-1:2025, the sixth edition. The recognition is partial: clause 6.9 (biological risk estimation) and the phrase "consumer products" in clause 6.5.11.3 are excluded, so biological risk in a US file is still estimated inside the ISO 14971 framework. The database entry states that FDA will accept declarations of conformity to the 2018 edition in premarket submissions until 1 July 2029.

The same list replaced 2-289 with 2-314 for ISO 10993-12:2021 including Amendment 1:2025, so the sample-preparation standard behind every extract is now recognized with its amendment.

At FDA's Biocompatibility Town Hall #1 on 9 September 2026, slide 11 stated that there are no changes to FDA's biocompatibility guidance related to ISO 10993-1:2025 at this time. That was a statement on a slide, not a guidance revision; the current guidance remains "Use of International Standard ISO 10993-1," issued September 2023.

Who it affects

Everyone filing in the US. The practical change is in how contact duration is counted: any contact within a day counts as a contact day, intermittent days are summed, and use beyond the labeled duration is treated as foreseeable misuse. One extra day can move a device from limited to prolonged contact and bring genotoxicity into scope.

What to do

You may still declare conformity to the 2018 edition until 1 July 2029, but new plans should be written to the 2025 structure and mapped to the exclusions. Re-check the contact-duration category of every open program under the 2025 counting rules before anything goes on study. Cite 2-314 for ISO 10993-12 with its amendment.

Sources

  1. FDA recognition 2-313, ISO 10993-1:2025 checked 20 Nov 2026
  2. Recognition List 066, 91 FR 54715 checked 20 Nov 2026
  3. FDA Biocompatibility Town Hall #1 slides, 9 September 2026 checked 20 Nov 2026
  4. FDA guidance, Use of International Standard ISO 10993-1 (September 2023) checked 20 Nov 2026

Tests and services this touches

May all we do lead to a safer world. Relentlessly Forward.