Arrow International (10 November) and Heraeus Medical Components (20 November) removed hydrophilic-coated guidewires after coatings swelled unevenly in saline and shed particles. Neither was an extractables finding.
- Who it affects
- Coated intravascular devices (guidewires, PICCs, stylets) and their biological evaluations
What changed
On 10 November 2026 FDA posted an early alert for Arrow International, a Teleflex subsidiary, which wrote to customers on 5 November 2026 about removing products that contain hydrophilic-coated spring wire guides: Arrow Marked nitinol and stainless-steel guidewires, pressure-injectable Arrowg+ard Blue Advance PICCs and JACCs with 80 cm or 130 cm coated guidewires, PICC kits preloaded with the Arrow VPS Precision Stylet, and other pressure-injectable PICCs. Teleflex confirmed that the coating may swell inconsistently when hydrated with saline, causing clumping and possible detachment, with particles able to enter the bloodstream and a risk of emboli.
On 20 November 2026 FDA posted a second early alert, for Heraeus Medical Components (Heraeus Medevio), which wrote to customers on 4 and 5 November about removing Navilyst, Spectrum and Arrow Marked vascular guidewires it manufactures. The description is the same: uneven swelling of the hydrophilic coating in saline, clumping most often on the distal coil, and a risk that the coating flakes or detaches.
As of 5 November 2026, neither company had reported serious injuries or deaths. Both alerts describe a coating-integrity failure, not an extractables or leachables finding.
Who it affects
Manufacturers and contract manufacturers of coated intravascular devices, and anyone whose biological evaluation treats a hydrophilic coating as a chemistry question only.
What to do
Put coating integrity and particulate data in the biological evaluation alongside the chemistry: hydration in the clinical fluid, flexure and simulated-use conditioning, and particulate counts by size, with acceptance criteria tied to the embolic risk. If the coating is supplied, confirm that the supplier's lot-release testing covers hydration behavior and that a change-notification clause is in the quality agreement. Two removals in one month for the same failure mode make this the question a reviewer will ask next.
Sources
- FDA early alert, Arrow International, 10 November 2026 checked 20 Nov 2026
- FDA early alert, Heraeus Medical Components, 20 November 2026 checked 20 Nov 2026