Safety alert: two hydrophilic-coating removals in one month put coating integrity in the evaluation

  • US United States (FDA)

Arrow International (10 November) and Heraeus Medical Components (20 November) removed hydrophilic-coated guidewires after coatings swelled unevenly in saline and shed particles. Neither was an extractables finding.

Who it affects
Coated intravascular devices (guidewires, PICCs, stylets) and their biological evaluations

What changed

On 10 November 2026 FDA posted an early alert for Arrow International, a Teleflex subsidiary, which wrote to customers on 5 November 2026 about removing products that contain hydrophilic-coated spring wire guides: Arrow Marked nitinol and stainless-steel guidewires, pressure-injectable Arrowg+ard Blue Advance PICCs and JACCs with 80 cm or 130 cm coated guidewires, PICC kits preloaded with the Arrow VPS Precision Stylet, and other pressure-injectable PICCs. Teleflex confirmed that the coating may swell inconsistently when hydrated with saline, causing clumping and possible detachment, with particles able to enter the bloodstream and a risk of emboli.

On 20 November 2026 FDA posted a second early alert, for Heraeus Medical Components (Heraeus Medevio), which wrote to customers on 4 and 5 November about removing Navilyst, Spectrum and Arrow Marked vascular guidewires it manufactures. The description is the same: uneven swelling of the hydrophilic coating in saline, clumping most often on the distal coil, and a risk that the coating flakes or detaches.

As of 5 November 2026, neither company had reported serious injuries or deaths. Both alerts describe a coating-integrity failure, not an extractables or leachables finding.

Who it affects

Manufacturers and contract manufacturers of coated intravascular devices, and anyone whose biological evaluation treats a hydrophilic coating as a chemistry question only.

What to do

Put coating integrity and particulate data in the biological evaluation alongside the chemistry: hydration in the clinical fluid, flexure and simulated-use conditioning, and particulate counts by size, with acceptance criteria tied to the embolic risk. If the coating is supplied, confirm that the supplier's lot-release testing covers hydration behavior and that a change-notification clause is in the quality agreement. Two removals in one month for the same failure mode make this the question a reviewer will ask next.

Sources

  1. FDA early alert, Arrow International, 10 November 2026 checked 20 Nov 2026
  2. FDA early alert, Heraeus Medical Components, 20 November 2026 checked 20 Nov 2026