All regulatory updates

Regulatory updateStandards and recognition

ISO 10993-3:2026 is published; FDA still recognizes the 2014 edition and TR 33

  • US United States (FDA)
Before

ISO 10993-3:2026 as a declaration of conformity (not recognized by FDA; not harmonised in the EU)

Cite now

ISO 10993-3:2014 — FDA recognition 2-228 (partial) with ISO/TR 10993-33:2015 — recognition 2-243 (partial); ISO 10993-3:2026 cited as supporting information with a justification

The fourth edition moves Part 3 toward evaluation and absorbs the TR 33 test guidance. For US declarations the recognized documents are still ISO 10993-3:2014 (2-228) and ISO/TR 10993-33:2015 (2-243).

Who it affects
Genotoxicity, carcinogenicity and reproductive toxicity endpoints; declarations of conformity in US submissions

What changed

ISO published ISO 10993-3:2026, the fourth edition, in August 2026, retitled "Evaluation of genotoxicity, carcinogenicity, reproductive toxicity and developmental toxicity." ISO describes it as a significant revision: it shifts emphasis from testing alone to a risk-based evaluation that considers intended use, exposure and biological interaction, states that chemical characterization and a toxicological risk assessment can inform whether testing is needed, and incorporates the genotoxicity test guidance that previously sat in ISO/TR 10993-33. ISO now lists the 2014 edition and TR 10993-33:2015 as withdrawn.

FDA's recognized consensus standards database has no entry for the 2026 edition as of 20 November 2026. The recognized documents remain ISO 10993-3:2014 (recognition 2-228, partial: clauses 4.1, 5.2.3, 5.3, the single-species sentence in 6.2, "or test material" in 7.2, and Annexes A and B are excluded) and ISO/TR 10993-33:2015 (recognition 2-243, partial). Decision (EU) 2026/2479 did not harmonise Part 3 either.

Who it affects

Programs with prolonged or long-term contact that now carry a genotoxicity endpoint under ISO 10993-1:2025, and anyone planning to argue a genotoxicity endpoint from chemistry rather than a test battery.

What to do

For US submissions, declare conformity to ISO 10993-3:2014 under 2-228 and to TR 33 under 2-243, and respect their exclusions. If the evaluation follows the 2026 approach, cite the 2026 edition as supporting information with a justification, not as a declaration of conformity. Watch for an FDA entry; when it appears this item will be marked superseded and linked forward.

Sources

  1. ISO 10993-3:2026 checked 20 Nov 2026
  2. FDA recognition 2-228, ISO 10993-3:2014 checked 20 Nov 2026
  3. FDA recognition 2-243, ISO/TR 10993-33:2015 checked 20 Nov 2026

Tests and services this touches

May all we do lead to a safer world. Relentlessly Forward.